1 what is an alarm – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 158

Advertising
background image

IM 11M13A01-04E

12-4

12.2 Displays and Measures to Take When Alarms are Gen-

erated

12.2.1

What is an Alarm?

When an alarm is generated, the alarm indication blinks in the display to notify of the

alarm (Figure 12.4).

Pressing the alarm indication displays a description of the alarm. Alarms include those
shown in Table 12.2.

Displayed alternately

AL-06

21.0%

F12.4E.EPS

Figure 12.4

Table 12.2 Types of Alarms and Reasons for Occurrence

T12.2E.EPS

Alarm

Type of alarm

Reason for occurrence

Oxygen concentration alarm

Alarm 1

concentration alarm

Occurs when a measured value exceed or falls below
the set alarm value (refer to Section 8.3, "Alarm Setting").

Alarm 6

Zero-point calibration coefficient alarm Generated when the zero correction factor is

out of the range of 100

Ϯ 30% in calibration

(refer to Section 9.1.3, "Compensation").

Alarm 7

Span-point calibration coefficient alarm Generated when the span correction factor is out

of the range of 0

Ϯ 18% in calibration (refer to Section

9.1.3, "Compensation").

Alarm 8

EMF stabilization time-up

Generated when the cell (sensor) voltage is not
stabilized even after the calibration time is up in
automatic and semiautomatic calibration.

Alarm 10

Cold junction temperature alarm

Occurs when an equipment internal temperature
exceeds 85

ЊC.

Alarm 11

Thermocouple voltage alarm

Generated when thermocouple voltage exceeds 42.1 mV
(about 1020

ЊC ) or falls below -5 mV (about -170ЊC).

Alarm 13

Battery low alarm

Internal battery needs replacement.

If an alarm is generated, such measures as turning off the heater power are not carried

out. The alarm is released when the cause for the alarm is eliminated.

However, Alarm 10 and/or Alarm 11 may be generated at the same time as Error-2
(heater temperature error).

In such a case, the measure taken for this error has priority.

If the converter power is turned off after an alarm is generated and restarted before the
cause of the alarm has been eliminated, the alarm will be generated again.

However, Alarms 6, 7, and 8 (alarms related to calibration) are not generated unless
calibration is executed.