IM 11M13A01-04E

7-10

7.5

Confirmation of Equipment Type Setting

This equipment can be used for both the Oxygen Analyzer and the Humidity Analyzer.
If you choose optional specification /HS at the time of purchase, the equipment is set for
the Humidity Analyzer.

Before setting the operating data, be sure to check that the desired model has been set.
Note that if the equipment type setting is changed after operating data are set, the
operating data that have been set are then initialized and the default settings remain. Set
the equipment type with parameter code F01. See Table 10.6, later in this manual.

CAUTION

Note that if the equipment type is changed, operation data that have already been set are
initialized (reverting to the default setting).

Table 7.5 Converter Type Setting Procedure

T7.3E.EPS

Switch operation

Display

Description

The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

Ͼ

A

01

F

01

F

01

0

0

0

0

Basic

panel

display

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

Ͼ

Ͼ

Display after the password has been entered.

Touch the [ ] key to switch to Group F. If an unwanted alphabetic character after

F has been entered, continuously touch the [>] key to return to the original.

Touch the [ENT] key for confirmation. If 0 (zero) is entered, the oxygen analyzer

is already set. If 1 (one) is entered, the humidity analyzer has been set.

Change the setting following the steps below.

Continuously touch the [ ] key, and the position of the digit will change from 1 to 0 to

1 to 0. Release the [ENT] key when 0 is displayed.

Touch the [ENT] key. The numeric value will flash.

Touch the [ENT] key again to stop the numeric value from flashing.

Touch the [ENT] key once again, and the display will change to the parameter code

Touch the [>] key together with the [ENT] key to return to the basic panel display.

(This is not required if you proceed to make another setting.)(The displayed numeric

characters indicate the measurement gas concentration.)