18 history of calibration time, 19 time, 20 software revision – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 125

Advertising
background image

IM 11M13A01-04E

10-7

10. Other Functions

21.0

p1

0.51

81.92

ez

e1

es

0

C

A

B

Zero-gas ratio = (B/A) x 100 (%) Correctable range: 100 – 30%
Span-gas ratio = (C/A) x 100 (%) Correctable range: 0 – 18%

Cell
electromotive
force, mV

Calibration curve
before correction

Previous
zero-gas data

Corrected calibration curve
(theoretical calibration curve)

Zero origin

Span origin

Span-gas
concentration

Oxygen concentration (vol%O

2

)

F10.2E.EPS

Figure 10.2

10.1.18 History of Calibration Time

The calibration-conducted dates and times for the past ten calibrations are stored in
memory.

10.1.19 Time

The current date and time are displayed. These are backed up by built-in batteries, so no
adjustment is required after the power is switched off. The following shows an example
of displaying June 21, 2000, 3:06 p.m.

Displayed alternately

00.06.21

15.06

F10.3E.EPS

Figure 10.3 Date-and-time Display

10.1.20 Software Revision

The revision (number) of the software installed is displayed.