Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 24

Advertising
background image

IM 11M13A01-04E

2-6

Installation: Flange mounting

Probe Mounting Angle: Horizontal to vertically downward.

Case: Aluminum alloy

Paint Color: Cover and case; Mint green (Munsell 5.6BG 3.3/2.9)

Finish: Polyurethane corrosion-resistance coating

Weight:

Insertion length of 0.4m: Approx. 15 kg (ANSI 150 4)

Insertion length of 1.0m: Approx. 17 kg (ANSI 150 4)

Insertion length of 1.5m: Approx. 19 kg (ANSI 150 4)

Insertion length of 2.0m: Approx. 21 kg (ANSI 150 4)

Functions

Display Function: Displays values of the measured oxygen concentration, etc Alarm,
Error Display: Displays alarms such as “AL-06” or errors such as “Err-01” when any
such status occurs.

Calibration Functions:

Auto-calibration; Requires the Autocalibration Unit. It calibrates automatically

at specified intervals.

Semi-auto Calibration; Requires the Autocalibration Unit. Input calibration
started by optical switch or contact, then it calibrates automatically afterwards.
Manual Calibration; Calibration with opening/closing the valve of calibration

gas in operation interactively with the optical switch.

Maintenance Functions:

Can operate updated data settings in daily operation and checking.

Display data settings, calibration data settings, test settings (current output loop

check, input/output contact check).

Setup Functions:

Initial settings suit for the plant conditions when installing the converter.
Current output data settings, alarm data settings, contact data settings, other
settings.

Display and setting content:

Display Related Items: Oxygen concentration (vol% O

2

), Output current value (mA), air

ratio, Cell temperature (

ЊC ), thermocouple reference junction temperature (ЊC ),

maximum/minimum/average oxygen concentration (vol% O

2

), cell e.m.f. (mV),

cell internal resistance (

⍀), cell condition (in four grades), heater on-time rate

(%), calibration record (ten times), time (year/month/day/hour/minute)

Calibration Setting Items: Span gas concentration (vol% O

2

), zero-gas concentration

(vol%O

2

), calibration mode (auto, semi-auto, manual), calibration type and

method (zero-span calibration, zero calibration only, span calibration only),

stabilization time (min.sec), calibration time (min.sec), calibration period (day/

hour), starting time (year/month/day/hour/minute)

Output Related Items: Analog output/output mode selection, output conditions when

warming-up/maintenance/calibrating/abnormal, 4 mA/20 mA point oxygen

concentration (vol% O

2

), time constant, preset values when warming-up/

maintenance/calibrating/abnormal, output preset values on abnormal

Alarm Related Items: Oxygen concentration high-alarm/high-high alarm limit values

(vol% O

2

), Oxygen concentration low-alarm/low-low alarm limit values (vol%

O

2

), Oxygen concentration alarm hysteresis (vol% O

2

), Oxygen concentra tion

alarm detection, alarm delay (seconds)

Advertising