2 piping for system 2, Figure 4.9 piping for system 2 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 54

Advertising
background image

IM 11M13A01-04E

4-4

4.2

Piping for System 2

Piping in System 2 is illustrated in Figure 4.9. In System 2, calibration is automated;
however, the piping is basically the same as that of System 1. Refer to Section 4.1.

Adjust secondary pressure of both the air set and the zero gas regulator so that these two
pressures are approximately the same. The flow rate of zero and span gases (normally
instrument air) are set by a individual needle valve. After installation and wiring, check
the calibration contact output (see Sec. 7.9.2), and adjust zero gas regulator and calibra-
tion gas needle valve so that zero gas flow is within the permitted range. Next check
span gas calibration contact output and adjust air set so that span gas flow is within the
permitted range.

~

100 to 240 V AC

Auto Calibration unit

Reference gas

Calibration gas (Zero)

Contact input
Analog output, contact output
Digital output (HART)

Air Set

Instrument air

Calibration gas
unit case

Pressure regulator

Zero gas cylinder

ZR202S Integrated type Explosionproof
Zirconia Oxygen Analyzer

F03.EPS

Span gas

Non-hazardous Area

Hazardous Area

Note:
The installation temperature limits range for integrated type analyzer is -20 to 55

؇

C.

*2

Figure 4.9 Piping for System 2