Caution, 3 alarm setting – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 96

Advertising
background image

IM 11M13A01-04E

8-8

In the example in Figure 8.2, the high-limit alarm point is set to 7.5% O

2

, the delayed

time is set to five seconds, and hysteresis is set to 2% O

2

.

Alarm output actions in this figure are expressed as follows:

(1) Although oxygen concentration measurement “A” has exceeded the high-limit alarm

setpoint, “A” falls lower than the high-limit alarm setpoint before the preset delayed
time of five seconds elapses. So, no alarm is issued.

(2) Oxygen concentration measurement “B” exceeds the high-limit alarm setpoint and

the delayed time has elapsed during that measurement. So, an alarm results.

(3) Although oxygen concentration measurement “C” has fallen lower than the hyster-

esis set value, that measurement exceeds the hysteresis set value before the preset
delayed time has elapsed. So, the alarm is not canceled.

(4) Oxygen concentration measurement “D” has fallen below the hysteresis set value and

the preset delayed time during measurement has elapsed, so the alarm is canceled.

8.4.3

Alarm Setting

Set the alarm setpoints following Table 8.7 listing parameter codes.

Table 8.7 Parameter Codes for Oxygen Concentration Alarms

Set item

Parameter code

Set value

Oxygen concentration

D01

0-100% O

2

high-high alarm setpoint

Oxygen concentration

D02

0-100% O

2

high-alarm setpoint

Oxygen concentration

D03

0-100% O

2

low-alarm setpoint

Oxygen concentration

D04

0-100% O

2

low-low-alarm setpoint

Oxygen concentration

D30

0-9.9% O

2

alarm hysteresis

Delayed alarm action

D33

0-255 seconds

Oxygen concentration

D41

0: Not detected

high-high alarm detection

1: Detected

Oxygen concentration

D42

0: Not detected

high-alarm detection

1: Detected

Oxygen concentration

D43

0: Not detected

low-alarm detection

1: Detected

Oxygen concentration

D44

0: Not detected

low-low-alarm detection

1: Detected

T8.7E.EPS

CAUTION

Even with alarms set, if “Disabled” has been set in the above alarm detection, no alarm
is issued. Be sure to set “Enabled” in the above alarm detection if you use alarm
features.