4 setting purging, Setting purging -18 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 101

Advertising
background image

<8. Detailed Data Setting>

8-18

IM 11M12A01-04E

8.7.3.2

Procedure

Use the parameter code table below to set fuel values.

Table 8.17

Setting Fuel Values

Set item

Parameter code Set value Engineering units

Amount of water vapor in

exhaust gas

F20

0 to 5

m

3

/kg (m

3

)

Theoretical amount of air

F21

1 to 20

m

3

/kg (m

3

)

X value

F22

0 to 19.99

Absolute humidity of the

atmosphere

F23

O to 1

kg/kg

8.7.3.3

Default Values

When the analyzer is delivered, or if data are initialized, parameter settings are by default, as shown in

Table 8.18.

Table 8.18

Default Settings of Fuel Values

Item

Default setting

Amount of water vapor in exhaust gas 1.00 m

3

/kg (m

3

)

Theoretical amount of air

1.00 m

3

/kg (m

3

)

X value

1.00

Absolute humidity of the atmosphere

0.1000 kg/kg

8.7.4 Setting Purging

Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration

gas for a given length of time before warm-up of the detector. This prevents cell breakage during

calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge time

has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100°C or below upon power up and the purge time is

set in the range of 1 to 60 minutes.

F8.5.ai

Displayed alternately

Figure 8.5

Display during Purging

8.7.4.1

Procedure

Use the parameter-code table below to set the purging time.
The allowable input ranges from 0 to 60 minutes.

Table 8.19

Purging Time

Set item

Parameter code

Set range

Units

Purging time

F15

0 to 60

minutes

8.7.4.2

Default Value

When the analyzer is delivered, or if data are initialized, purging time is set to 0 minutes.