1.1.3 System 3

This example, System 3, represents typical applications in large boilers and heating furnaces, where

is a need to monitor and control oxygen concentration. The reference gas and calibration-time span

gas are (clean, dry) instrument air. Zero gas is supplied from a gas cylinder.
System 3 uses the ZR20H automatic calibration unit, with auto-switching of the calibration gas.
A “combustible gas detected” contact input turns off power to the heater. There’s also contact output

from the converter that can be used to operate a purge gas valve to supply air to the sensor.

~

100 to 240 V AC

Automatic calibration unit

Reference gas

Calibration gas (Zero)

Contact input

Analog output, contact output

Digital output (HART)

Air Set

Instrument air

Calibration gas unit case

Calibration

gas pressure

regulator

Zero gas cylinder

ZR202G Integrated type Zirconia Oxygen Analyzer
with automatic calibration unit (ZR202G-

-

-

-A-

-

-

-

-A)

F1.3E.ai

Span gas

ZR20H

*3

Note:

The installation temperature limits range for integrated type analyzer is -20 to 55°C.

*2

*1

*

1

Shield cable:

Use shielded signal cables, and connect the shields to the FG terminal of the converter.

*2

Select the desired probe from the Probe Configuration table on page 1-4.

*3

When a zirconia oxygen analyzer is used, 100% N

2

gas cannot be used as the zero gas.

Use approx. 1 vol% O

2

gas (N

2

-based).

Figure 1.3

Example of System 3