<5. Wiring>

5-3

IM 11M12A01-04E

5.1.3 Mounting of Cable Gland

For each wiring inlet connection of the equipment, mount the conduit appropriate for the screw size or

a cable gland.

F5.3E.ai

25

4-G1/2,2-1/2NPT etc.

Cable connection port

Cable gland

Rc1/4 or 1/4NPT

Reference gas inlet

Rc1/4 or 1/4NPT

Calibration gas inlet

Figure 5.3

Cable Gland Mounting

5.2 Wiring for Analog Output

This wiring is for transmitting 4 to 20mA DC output signals to a device, e.g. recorder. Maintain the load

resistance including the wiring resistance of 550Ω or less.

+

-

AO(+)
AO(

-

)

FG

F5.4E.ai

Shielded cables

Analyzer

Receiver

Figure 5.4

Wiring for Analog Output

5.2.1 Cable Specifications

Use a 2-core shielded cable for wiring.

5.2.2 Wiring Procedure

(1) M4 screws are used for the terminals. Use crimp-on terminals appropriate for M4 terminal

screws for cable connections. Ensure that the cable shield is connected to the FG terminal of the

equipment.

(2) Be sure to connect (+) and (-) polarities correctly.

CAUTION

• Before opening the cover, loosen the lock screw. If the screw is not loosened first, the cover

will be improperly engaged to the body, and the terminal box will require replacement. When

opening and closing the cover, remove any sand particles or dust to avoid gouging the thread.

• After screwing the cover on the equipment body, secure it with the lock screw.