<3. Installation>

3-10

IM 11M12A01-04E

3.5 Insulation Resistance Test

Even if the testing voltage is not so great that it causes dielectric breakdown, testing may cause

deterioration in insulation and a possible safety hazard. Therefore, conduct this test only when it is

necessary.
The applied voltage for this test shall be 500 V DC or less. The voltage shall be applied for as short a

time as practicable to confirm that insulation resistance is 20 MΩ or more.
Remove wiring from the converter and the detector.

1. Remove the jumper plate located between terminal G and the protective grounding terminal.

2. Connect crossover wiring between L and N.

3. Connect an insulation resistance tester (with its power OFF). Connect (+) terminal to the

crossover wiring, and (-) terminal to ground.

4. Turn the insulation resistance tester ON and measure the insulation resistance.

5. After testing, remove the tester and connect a 100 kΩ resistance between the crossover wiring

and ground, to discharge.

6. Testing between the heater terminal and ground, contact output terminal and ground, analog

output/input terminal and the ground can be conducted in the same manner.

7. Although contact input terminals are isolated, insulation resistance test cannot be conducted

because the breakdown voltage of the surge-preventing arrester between the terminal and

ground is low.

8. After conducting all the tests, replace the jumper plate as it was.

1

DI-1

2

DI-2

3

DI-C

4

DO-1

5

DO-1

6

DO-2

7

DO-2

8

FG

9

AO

(+)

10

AO

(-)

11

L

12

N

13

G

14

FG

F3.17E.ai

Remove

jumper

plate

Insulation

resistance

tester

Insulation

resistance

tester

Crossover wiring

Crossover wiring

Contact input 1

Contact input 2

+

-

+

-

Insulation

resistance

tester

+

-