2 setting output contact, Setting output contact -9, Warning – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 92

Advertising
background image

<8. Detailed Data Setting>

8-9

IM 11M12A01-04E

8.5.2 Setting Output Contact

Set the output contacts following Table 8.10.

Table 8.10

Parameter Codes for Output Contact Setting

Set item

Parameter

code

Set value

Output contact 1
Operation

E10

0 Operated in closed status. (Normally de-energized)
1 Operated when open. (Normally energized)

(Note 1)

Error

E20

0 Not operated if an error occurs.
1 Operated if an error occurs.

High-high alarm

E21

0 Not operated if a high-high alarm occurs.
1 Operated if a high-high alarm occurs.

(Note 2)

High alarm

E22

0 Not operated if a high alarm occurs.
1 Operated if a high alarm occurs.

(Note 2)

Low alarm

E23

0 Not operated if a low alarm occurs.
1 Operated if a low alarm occurs.

(Note 2)

Low-low alarm

E24

0 Not operated if a low-low alarm occurs.
1 Operated if a low-low alarm occurs.

(Note 2)

Maintenance

E25

0 Not operated during maintenance.
1 Operated during maintenance (see Section 8.3.1).

Calibration

E26

0 Not operated during calibration.
1 Operated during calibration (see Section 8.3.1).

Measurement range

change

E27

0 Not operated when changing ranges.
1 Operated when changing ranges.

(Note 3)

Warm-up

E28

0 Not operated during warming up.
1 Operated during warming up.

Calibration gas

pressure decrease

E29

0 Not operated while a calibration gas pressure decrease contact is

being closed.

1 Operated while a calibration gas pressure decrease contact is being

closed.

(Note 4)

Unburnt gas

detection

E32

0 Not operated while a unburnt gas detection contact is being closed.
1 Operated while a unburnt gas detection contact is being closed.

(Note 5)

Note 1: Output contact 2 remains closed.

Note 2: The oxygen concentration alarm must be preset (see Section 8.4).

Note 3: Range change answer-back signal. For this action, the range change must be preset during the setting of input contacts

(see Section 8.6).

Note 4: Calibration gas pressure decrease answer-back signal. Calibration gas pressure decrease must be selected beforehand

during the setting of input contacts.

Note 5: Non-combusted gas detection answer-back signals. “Non-combusted gas” detection must be selected during the setting of

input contacts.

WARNING

• Output contact 2 is linked to the detector’s heater power safety switch. As such, if output

contact 2 is on, the heater power stops and an Error 1 (cell voltage abnormal) or Error 2 (heater

temperature abnormal) occurs.

Advertising