<10. Other Functions>

10-15

IM 11M12A01-04E

Equipment Setup and Others in Group F

Code

Item

Tuning

Engineering unit

Default setting

F01

Equipment setup

0 Oxygen analyzer

Oxygen analyzer

1 Humidity analyzer

F02

Selection of measurement

gas

0 Wet

Wet

1 Dry

F04

Selection of temperature

units

0 degree C

degree C

1 degree F

F05

Selection of pressure units 0 kPa

kPa

1 psi

F08 Selection of display items

0 Oxygen concentration

Oxygen

concentration

1 Amount of moisture quantity
2 Mixed ratio
3 Item selected with analog output

F10

Date

YY.MM.DD/HH.MM

F11 Period over which average

values are calculated

1 to 255 hours

Hours

One hour

F12 Period over which max.

and min. values are

monitored

1 to 255 hours

Hours

24 hours

F15

Purging time

0 to 60 minutes

Minutes

0 minute

F20 Amount of water vapor in

exhaust gas

0 to 5

m

3

/kg (m

3

)

1.0 m

3

/kg (m

3

)

F21 Theoretical amount of air

0 to 20

m

3

/kg (m

3

)

1.0 m

3

/kg (m

3

)

F22 X value

0 to 19.99

1

F23 Absolute humidity of the

atmosphere

0 to 1

kg/kg

0.1 kg/kg

F30 Initializing all data
F31 Initializing data in group A
F32 Initializing data in group B
F33 Initializing data in group C
F34 Initializing data in group D
F35 Initializing data in group E
F36 Initializing data in group F

Inspection-related Items in Group G

Code

Item

Tuning

Engineering unit

Default setting

G01 mA-output loop

4 to 20

mA

4 mA

G11

Output contact 1

0 Open

Open

1 Closed

G12

Output contact 2

0 Open

Open

1 Closed

G15 Automatic calibration

solenoid valve (zero)

0 Off

Off

1 On

G16 Automatic calibration

solenoid valve (span)

0 Off

Off

1 On

G21

Input1 contact

0 Open
1 Closed

G22

Input2 contact

0 Open
1 Closed

G30 Reset