Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 105

Advertising
background image

<9. Calibration>

9-4

IIM 11M12A01-04E

9.1.4 Characteristic Data from a Sensor Measured During Calibration

During calibration, calibration data and sensor status data (listed below) are acquired. However, if

the calibration is not properly conducted (an error occurs in automatic or semi-automatic calibration),

these data are not collected in the current calibration.
These data can be observed using parameter codes

「A20」 to 「A22」, and 「A50」 to 「A79」. For an

explanation and the operating procedures of individual data, consult Section 10.1, “Detailed Display.”

(1) Record of span correction ratio

Recorded the past ten span correction ratios.

(2) Record of zero correction ratio

Recorded the past ten zero correction ratios.

(3) Response time

You can monitor the response time provided that a two-point calibration has been done in semi-

automatic or automatic calibration.

(4) Cell’s internal resistance

The cell’s internal resistance gradually increases as the cell (sensor) deteriorates. You can

monitor the values measured during the latest calibration. However, these values include the

cell’s internal resistance and other wiring connection resistance. So, the cell’s degrading cannot

be estimated from these values only.

When only a span calibration has been made, these values will not be measured, and previously

measured values will remain.

(5) Robustness of a cell

The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed

in a number on four levels.

Advertising