<7. Startup>

7-4

IM 11M12A01-04E

CAUTION

1. Be sure to put the equipment case cover back on. If this is not done, the infrared switch will not

reflect the infrared light-waves, and a “dSPErr” error will be issued.

2. Before placing the equipment in operation, be sure to wipe off any moisture or dust on the glass

surface if it is wet or dirty. Also make sure your fingers are clean and dry before touching the

glass surface of the switch.

3. If the infrared switches are exposed to direct sunlight, they may not operate correctly. In such a

case, change position of the display or install a sun cover.

7.4.2 Display Configuration

The parameter codes provided for the equipment are used to control the equipment display panels

(see below). By selecting appropriate parameter codes, you can conduct calibration and set operation

parameters. Figure 7.4 shows the configuration of display items. The parameter codes are listed in

groups of seven; which are briefly described in Table 7.2.
To enter parameters, you first need to enter the password, refer to See 7.4.3.
Touch the [ >] key and [ ENT] key at same time to revert to the main screen.

F7.4E.ai

Group G setup display

Basic panel display

Password entry display

Group A setup display

Group B setup display

Group C setup display

Parameter code

selection display

Group D setup display

Group E setup display

Group F setup display

%

Figure 7.4

Display Configuration