<11.Inspection and Maintenance>

11-10

IM 11M12A01-04E

11.3 Replacement of Flowmeter for ZR20H Automatic

Calibration Unit

(1) Remove pipe holding piping fitting

(2) Remove bolts holding flowmeter, and replace it. A white back plate (to make the float easy to

see) is attached. The end of the pin holding down the back plate must be on the bracket side.

(3) Replace piping, and fix M6 bolts between brackets. *1

*1 : When disassembling and reassembling, mark original positions, and tighten an extra 5-10°

when reassembling. After tightening, do a liquid leakage test.

Vertical mounting

Horizontal mounting

C

C

B

B

C'

B'

A'

A

A

Zr20h_g0.ai

Connect piping pairs A-A9, B-B9, C-C9, D-D9

Fixing screw pairs

Figure 11.9

Fixing Flowmeter