B0 3 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 81

Advertising
background image

<7. Startup>

7-16

IM 11M12A01-04E

Table 7.12

Calibration Setup Procedure

Switch operation

Display

Description

>

ENT

A

01

Display after the password has been entered.

>

ENT

b

01

Set the zero gas concentration. Switch the parameter code to B01.
Here, set 0.98%.

>

ENT

0

01.00

%

Touch the [ENT] key to display the currently set value.

>

ENT

00

1

.00

%

Touch the [>] key to move the position of the digit that is flashing to 1.

>

ENT

00

0

.00

%

Touch the [∧] key to change to 0.

>

ENT

000.

0

0

%

Touch the [>] key to move the position of the digit that is flashing to the
right one digit.

>

ENT

000.

9

0

%

Touch the [∧] key to change the numeric value to 9.

>

ENT

000.9

0

%

Touch the [>] key to move the position of the digit that is flashing to the
right one digit.

>

ENT

000.9

8

%

Touch the [∧] key to change the numeric value to 8.

>

ENT

000.98

%

Touch the [ENT] key to have all the digits flash.

>

ENT

000.98

%

Touch the [ENT] key again to stop the flashing.

>

ENT

b

01

Touch the [ENT] key once again to switch to the parameter code selection
display.

Set the span gas concentration by above procedure, set 21 %.

>

ENT

b0

3

Next, set the calibration mode. Switch the parameter code to B03.

>

ENT

0

Touch the [ENT] key to display the currently set value.
If it is 0, you can leave it as is. If it is other than 0, change it to 0 (zero).

>

ENT

0

Touch the [ENT] key. The numeric value will flash.

>

ENT

0

Touch the [ENT] key again to stop the flashing.

>

ENT

b

03

Touch the [ENT] key once again to switch to the parameter code selection
display.

>

ENT

Basic panel

display

Touch the [>] key together with the [ENT] key to return to the basic panel
display. (This is not required if you proceed to make another setting.)
(The displayed numeric characters indicate the measurement gas
concentration.)

The symbol [

] indicates that the key is being touched. Light characters indicates that the digits are flashing.

Advertising