4 reset, Reset -9, Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 118

Advertising
background image

<10. Other Functions>

10-9

IM 11M12A01-04E

10.4 Reset

Resetting enables the equipment to restart. If the equipment is reset, the power is turned off and

then back on. In practical use, the power remains on, and the equipment is restarted under program

control. Resetting will be possible in the following conditions:

(1) Error 1

if the cell voltage is defective

(2) Error 2

if a temperature alarm occurs

(3) Error 3

if the A/D converter is defective

(4) Error 4

if an EEPROM write error occurs

For details on error occurrence, consult Chapter 12, “Troubleshooting”, later in this manual.
If any of the above problems occurs, the equipment turns off the power to the detector heater. To

cancel the error, reset the equipment following the steps below, or turn the power off and then back

on.

CAUTION

• Make sure that before resetting or restarting the power that there is no problem with the

equipment.

• If a problem arises again after the resetting, turn the power off and troubleshoot the problem by

consulting the Troubleshooting chapter later in this manual. When there is no error, the Basic

panel display will appear.