4 selecting parameter codes, 5 changing set values, Selecting parameter codes -6 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 71: Changing set values -6, A0 1

Advertising
background image

<7. Startup>

7-6

IM 11M12A01-04E

7.4.4 Selecting Parameter Codes

Switch operation

Display

Description

>

ENT

A01

Password has been entered and the parameter code selection display has
appeared.
Character A is flashing, indicating that character A can be changed.

>

ENT

A

0

1

If you touch the [>] key once, the position of the digit that is flashing will move to
the right.
This allows you to change 0.

>

ENT

A0

1

Touch the [>] key again to move the position of the digit that is flashing to the right
one more digit. This enables you to change numeric character 1.

>

ENT

A

01

Touch the [>] key again to return the position of the digit that is flashing to A.
Continuously touch the [>] key, and the position of the digit that is flashing will
move continuously to the right.

>

ENT

b

01

If you touch the [∧] key once, character A will change to B.

>

ENT

C

01

Touch the [∧] key once to change to C.

>

ENT

d

01

Continuously touch the [∧] key, and the value of the digit that is flashing will
increase continuously, from D to E to F to G to A. Numeric values will change from
0 to 1 to 2 to 3 … to 8 to 9 and back to 0. However, numbers that are not present
in the parameter codes will be skipped. Each digit is changed independently. Even
though a low-order digit changes from 9 to 0, a high-order digit will not be carried.

>

ENT

Set

Value

After you select the desired character, touch the [ENT] key. The set data will be
displayed.

The symbol [

] indicates that the key is being touched. Light characters indicates that the digits are flashing.

7.4.5 Changing Set Values

(1) Selecting numeric values from among preset values

Switch operation

Display

Description

>

ENT

0

The set value is displayed after the parameter code selection. An example of how
to select either 0, 1, or 2 as the set value is given below.
(The currently set value is 0.)

>

ENT

1

Touch the [∧] key once to change the current value from 0 to 1.

>

ENT

2

Touch the [∧] key again to change to the numeric value 2.

>

ENT

0

If you touch the [∧] key again, the numeric value will return to 0.
Continuously touch the [∧] key, and the numeric values will change continuously.

>

ENT

C

01

Display the desired numeric value and touch the [ENT] key.
The display will then return to the parameter code selection

Advertising