<8. Detailed Data Setting>

8-8

IM 11M12A01-04E

8.4.4 Default Values

When the analyzer is delivered, or if data are initialized, the default alarm set values are as shown in

Table 8.8.

Table 8.8

Alarm Setting Default Values

Set item

Set value

High-high alarm setpoint

100% O

2

High alarm setpoint

100% O

2

Low alarm setpoint

0% O

2

Low-low alarm setpoint

0% O

2

Alarm hysteresis

0.1% O

2

Delayed alarm action

3 seconds

High-high alarm detection

Not detected

High alarm detection

Not detected

Low alarm detection

Not detected

Low-low alarm detection

Not detected

8.5 Output Contact Setup

8.5.1 Output Contact

Mechanical relays provide contact outputs. Be sure to observe relay contact ratings. (For details, see

Section 2.1, "General Specifications.") The following sets forth the operation mode of each contact

output. Output contact 1 you can select open or closed contact when the contact is "operated". For

output contact 2, contact is closed. The relay for output contact 1 is energized when its contacts are

closed and vice versa. Accordingly, when no power is supplied to the equipment, those contacts

remain open. In addition, the relay for output contact 2 is energized when the corresponding contact is

open and de-energized when that contact is closed.

Table 8.9

Setting Output Contacts

Operating state

When no power is applied to this equipment

Output contact 1

Open (de-energized) or closed

(energized) selectable.

Open

Output contact 2

Closed (de-energized) only.

Closed