<12. Troubleshooting>

12-7

IM 11M12A01-04E

In addition, if an error occurs in calibration even after the cell assembly is replaced, the influence of

sample gas flow may be suspected. Do not let the sample gas flow toward the tip of the detector

probe, for example, by changing the mounting position of the detector.

12.2.2.5

Alarm 10: Cold Junction Temperature Alarm

The equipment incorporates a temperature sensor. An alarm is issued when the sensor temperature

exceeds 85°C. If internal temperature of this equipment exceeds 85°C, the electronics may fail.

<Search for cause of failure and Take measures>

This equipment can be used at ambient temperatures up to 55°C. If the ambient temperatures may

exceed the limits, take appropriate measures such as applying heat insulating material to the furnace

walls, and adding a sun shield to keep out direct sunlight.
If this alarm occurs even when the ambient temperature is under 55°C, the electronics may be

defective. Contact your local Yokogawa service or sales representative.

12.2.2.6

Alarm 11: Thermocouple Voltage Alarm

This alarm is generated when the emf (voltage) of thermocouple falls below -5 mV (about -170°C) or

exceeds 42.1 mV (about 1020°C).

(1) A failure of the thermocouple at the detector occurred.

(2) A failure of the electrical circuits occurred.

<Search for cause of failure and Take measures>

(1) Turn off the power to the analyzer.

(2) Remove the probe from the analyzer. Also remove all the connectors between the converter

and probe. Measure the resistance of the heater wire (yellow wire) from the probe as indicated

in Figure 12.5. The heater assembly is normal if the resistance is lower than about 90Ω. If the

resistance is higher than that value, the heater assembly may be defective. In this case, replace

the heater assembly (refer to Section 11.1.3, “Replacement of the Heater Assembly”).

F12.2E.ai

Multimeter

(Ω)

Heater wire

Figure 12.5

(3) Next, check the resistance of the thermocouple from the probe. Use a multimeter to measure the

thermocouple resistance between terminal 3 (red cable connected) and terminal 4 (white cable

connected) as indicated in Figure 12.6.

The thermocouple is normal if the resistance is 5Ω or less. If the value is higher than 5Ω, the

thermocouple wire may be broken or about to break. In this case, replace the heater assembly (refer

to Section 11.1.3, “Replacement of the Heater Assembly”).