<11.Inspection and Maintenance>

11-8

IM 11M12A01-04E

11.2 Inspection and Maintenance of the Converter

The converter does not require routine inspection and maintenance. If the converter does not work

properly, in most cases it probably comes from problems or other causes.

11.2.1 Replacing Fuses

This equipment incorporates a fuse. If the fuse blows out, turn off the equipment power and replace it

in the following procedure.

CAUTION

If a replaced fuse blows out immediately, there may be a problem in the circuit. Check the circuit

carefully to find out why the fuse has blown.
Before removing the electronics, touch the grounded metal part to discharge any static electricity.

(1) Remove the display cover (Figure 11.5).
(2) Remove the three screws that are located toward you, among the four screws shown in Figure

11.6. Loosen the remaining one.

(3) Move the electronics up to remove it.

F11.5E.ai

Cover of Display

F11.6E.ai

SCREW

Figure 11.5

Cover of Display

Figure 11.6

Location of Screw

(4) Disconnect the three connectors from the printed-circuit board, as shown in Figure 11.7, by

holding the connector housing. Do not pull the leadwire out to remove the connectors, otherwise,

disconnection may result.

(5) Remove the electronics completely to gain access to the fuse on the bottom of the equipment

case (Figure 11.8).

(6) Replace the fuse with a new one.

F11.7E.ai

F11.8E.ai

Figure 11.7

Locations of Connectors

Figure 11.8

Location of Fuse