Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 121

Advertising
background image

<10. Other Functions>

10-12

IM 11M12A01-04E

Calibration-related Items in Group B

Code

Item

Tuning

Engineering unit

Default setting

B01 Zero gas concentration

0.3 to 100

% O

2

1% O

2

B02 Span gas concentration

4.5 to 100

% O

2

21% O

2

B03 Calibration mode

0 Manual calibration

Manual calibration

1 Semi-automatic and manual

calibration

2 Automatic, semi-automatic, and

manual calibration

B04 Hold (Output stabilization)

time

0 minutes, 0 seconds to 60 minutes,

59 seconds

MM.SS

10 minutes, 0

seconds

B05 Calibration time

0 minutes, 0 seconds to 60 minutes,

59 seconds

MM.SS

10 minutes, 0

seconds

B06 Calibration interval

0 days 0 hours to 255 days 23 hours

DD.HH

30 days, 0 hours

B07 Calibration start date and

time

YY.MM.DD/HH.MM 00.01.01.00.00

B08 Calibration procedure

0 Zero and span

Zero and span

1 Span only
2 Zero only

B09 Calibration concentration

measurement

Display only

% O

2

B10 Manual calibration

implementation

B11 Semi-automatic calibration

implementation