2 installation of za8f flow setting unit, 1 installation location, Installation of za8f flow setting unit -5 3.2.1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 44: Installation location -5

Advertising
background image

<3. Installation>

3-5

IM 11M12A01-04E

<Detector with a probe protector (Model ZO21R-L-- *B for dust wear protect>

The detector is used with a probe protector to prevent the sensor from being worn by dust particles

when there is a high concentration of dust and gas flow exceeds 10 m/sec (fine-carbon boiler or fluid-

bed furnace).

(1) Put the gasket that is provided by user between the flanges, and mount the probe protector in

the probe insertion hole. The probe protector should be installed so that the notch comes to the

downstream of the sample gas flow.

(2) Make sure that the sensor assembly mounting screws (four bolts) at the probe tip are not loose.

(3) Where the detector is mounted horizontally, the calibration/reference gas inlet should face

downward.

Direction of the sample gas flow

Mount the protector so that the notch

is on the downstream side of gas flow.

F3104E.ai

1050,1550, 2050

Gasket (t1.5)

Unit: mm

Detector top

Calibration gas inlet

Reference gas inlet

Figure 3.4

Mounting of detector with a probe protector (Dust wear protect)

NOTE

When the probe protector is used in the ZR202G with pressure compensation (-P), instrument air

leaking from the probe protector may affect the measured value.

3.2 Installation of ZA8F Flow Setting Unit

3.2.1 Installation Location

The following should be taken into consideration:

(1) Easy access to the unit for checking and maintenance work.

(2) Near to the detector and the converter

(3) No corrosive gas.

(4) An ambient temperature of not more than 55°C and little changes of temperature.

(5) No vibration.

(6) Little exposure to rays of the sun or rain.

Advertising