Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 24

Advertising
background image

<2. Specifications>

2-5

IM 11M12A01-04E

Converter Output:

One mA analog output point (4 to 20 mA DC (maximum load resistance of

550Ω)) with mA digital output point (HART) (minimum load resistance of

250Ω).
Range; Any setting between 0 to 5 through 0 to 100 vol% O

2

in 1 vol% O

2

,

and partial range is available (Maximum range value/minimum

range value 1.3 or more)

For the log output, the minimum range value is fixed at 0.1 vol% O

2

.

4 to 20 mA DC linear or log can be selected. Input/output isolation.

Output damping; 0 to 255 seconds.

Hold/non-hold selection, preset value setting possible with hold.

Contact Output:

Two points, contact capacity 30V DC 3A, 250V AC 3A (resistive load)

One of the output points can be selected to ether normally energized or nor-

mally de-energized status.

Delayed functions (0 to 255 seconds) and hysteresis function (0 to 9.9 vol%

O

2

) can be added to high/low alarms.

The following functions are programmable for contact outputs.

(1) Abnormal, (2) High-high alarm, (3) High alarm, (4) Low-low alarm, (5)

Low alarm, (6) Maintenance, (7) Calibration, (8) Range switching answer-

back, (9) Warm-up, (10) Calibration gas pressure decrease (answer-back of

contact input), (11) Flameout gas detection (answer-back of contact input).

Contact Input:

Two points, voltage-free contacts

The following functions are programmable for contact inputs:

(1) Calibration gas pressure decrease alarm, (2) Range switching (switched

range is fixed), (3) External calibration start, (4) Process alarm (if this signal

is received, the heater power turns off)

Self-diagnosis:

Abnormal cell, abnormal cell temperature (low/high), abnormal calibration,

A/D converter abnormal, digital circuit abnormal

Calibration:

Method; zero/span calibration

Calibration mode; Auto, semi-auto and manual (All are operated using

optical switches). Either zero or span can be skipped.

Zero calibration gas concentration setting range; 0.3 to 100 vol% O

2

(in 0.01

vol% O

2

).

Span calibration gas concentration setting range; 4.5 to 100 vol% O

2

(in 0.01

vol% O

2

).

Use nitrogen-balanced mixed gas containing 0 to 10 vol% O

2

scale of

oxygen for standard zero gas and 80 to 100 vol% O

2

scale of oxygen for

standard span gas.

Calibration period; date/time setting: maximum 255 days