<12. Troubleshooting>

12-3

IM 11M12A01-04E

(3) Next, check the resistance of the thermocouple from the probe. Use a multimeter to measure the

thermocouple resistance between terminal 3 (red cable connected) and terminal 4 (white cable

connected) as indicated in Figure 12.3.

The thermocouple is normal if the resistance is 5Ω or less. If the value is higher than 5Ω, the

thermocouple wire may be broken or about to break. In this case, replace the heater assembly (refer

to Section 11.1.3, “Replacement of the Heater Assembly”).

CAUTION

Measure the thermocouple resistance value after the difference between the probe tip temperature

and the ambient temperature decreases to 50°C or less. If the thermocouple voltage is large,

accurate measurement cannot be achieved.

F12.3E.ai

1

2

3

4

YEL

GRN

RED

WHT

Thermocouple

Multimeter
(Ω)

Figure 12.3

(4) If the inspection indicates that the thermocouple is normal, the electronics may be defective.

Consult your local Yokogawa service or sales representative.

12.1.2.3

Error-3: A/D Converter Failure/Error-4: Writing-to-memory Failure

A/D Converter Failure

It is suspected that a failure has occurred in the A/D converter mounted in the converter

electronics.

Writing-to-memory Failure

It is suspected that a failure has occurred in an operation writing to the memory (EEPROM)

mounted in the converter electronics.

<Locating the failure and take measures>

Turn off the power to the converter once and then restart the converter. If the converter operates

normally after restarting, an error might have occurred due to a temporary drop in the voltage (falling

below 85 V, the least amount of voltage required to operate the converter) or a malfunction of the

electronics affected by noise. Check whether or not there is a failure in the power supply system or

whether the converter and detector are securely grounded.
If the error occurs again after restarting, a failure in the electronics is suspected. Consult the service

personnel at Yokogawa Electric Corporation.