Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 111

Advertising
background image

<10. Other Functions>

10-2

IM 11M12A01-04E

Table 10.1 Parameter Codes for Detailed Operation Data

Code

Item

Engineering

unit

Code

Item

Engineering

unit

A00 Selection

of display

items

0 Oxygen concentration

A50 Span correction ratio 0

%

1 Oxygen analyzer (0.0)

A51 Span correction ratio 1

%

2 Oxygen analyzer (0.0)

A52 Span correction ratio 2

%

3 Analog output selected

A53 Span correction ratio 3

%

A01 Oxygen concentration

% O

2

A54 Span correction ratio 4

%

A02

A55 Span correction ratio 5

%

A03

A56 Span correction ratio 6

%

A04

A57 Span correction ratio 7

%

A05

A58 Span correction ratio 8

%

A06 Air ratio

A59 Span correction ratio 9

%

A07 Cell temperature

°C

A60 Zero correction ratio 0

%

A08 Cold junction temperature

°C

A61 Zero correction ratio 1

%

A09 Meas. gas temperature

°C

A62 Zero correction ratio 2

%

A10 Amount of water vapor in exhaust gas

%

A63 Zero correction ratio 3

%

A11 Cell voltage

mV

A64 Zero correction ratio 4

%

A12 TC voltage

mV

A65 Zero correction ratio 5

%

A15 Cold junction voltage

mV

A66 Zero correction ratio 6

%

A16 Output current

mA

A67 Zero correction ratio 7

%

A20 Cell response time

Seconds

A68 Zero correction ratio 8

%

A21 Cell internal resistance

A69 Zero correction ratio 9

%

A22 Cell robustness

A70 Calibration history 0

YY.MM.

DD/HH.MM

A23 Heater on-time ratio

%

A71 Calibration history 1

YY.MM.

DD/HH.MM

A24 Oxygen concentration (with time

constant)

% O

2

A72 Calibration history 2

YY.MM.

DD/HH.MM

A25

A73 Calibration history 3

YY.MM.

DD/HH.MM

A26

A74 Calibration history 4

YY.MM.

DD/HH.MM

A30 Max. oxygen concentration

% O

2

A75 Calibration history 5

YY.MM.

DD/HH.MM

A31 Occurrence of maximum oxygen

concentration

YY.MM.

DD/HH.MM

A76 Calibration history 6

YY.MM.

DD/HH.MM

A32 Min. oxygen concentration

% O

2

A77 Calibration history 7

YY.MM.

DD/HH.MM

A33 Occurrence of minimum oxygen

concentration

YY.MM.

DD/HH.MM

A78 Calibration history 8

YY.MM.

DD/HH.MM

A34 Average oxygen concentration

% O

2

A79 Calibration history 9

YY.MM.

DD/HH.MM

A35

A80 Time

YY.MM.

DD/HH.MM

A36

A90 Software revision

Note: The blank parameter codes above are not used in the oxygen analyzer.