<8. Detailed Data Setting>

8-7

IM 11M12A01-04E

In the example in Figure 8.2, the high alarm point is set to 7.5% O

2

, the delayed time is set to five

seconds, and hysteresis is set to 2% O

2

.

Alarm output actions in this figure are expressed as follows:

(1) Although oxygen concentration measurement “A” has exceeded the high alarm setpoint, “A” falls

lower than the high alarm setpoint before the preset delayed time of five seconds elapses. So,

no alarm is issued.

(2) Oxygen concentration measurement “B” exceeds the high alarm setpoint and the delayed time

has elapsed during that measurement. So, an alarm results.

(3) Although oxygen concentration measurement “C” has fallen lower than the hysteresis set value,

that measurement exceeds the hysteresis set value before the preset delayed time has elapsed.

So, the alarm is not canceled.

(4) Oxygen concentration measurement “D” has fallen below the hysteresis set value and the preset

delayed time during measurement has elapsed, so the alarm is canceled.

8.4.3 Alarm Setting

Set the alarm setpoints following Table 8.7 listing parameter codes.

Table 8.7

Parameter Codes for Oxygen Concentration Alarms

Set item

Parameter code

Set value

Oxygen concentration

high-high alarm setpoint

D01

0-100% O

2

Oxygen concentration

high alarm setpoint

D02

0-100% O

2

Oxygen concentration

low alarm setpoint

D03

0-100% O

2

Oxygen concentration

low-low alarm setpoint

D04

0-100% O

2

Oxygen concentration

alarm hysteresis

D30

0-9.9% O

2

Delayed alarm action

D33

0-255 seconds

Oxygen concentration

high-high alarm detection

D41

0 Not detected
1 Detected

Oxygen concentration

high alarm detection

D42

0 Not detected
1 Detected

Oxygen concentration

low alarm detection

D43

0 Not detected
1 Detected

Oxygen concentration

low-low alarm detection

D44

0 Not detected
1 Detected

CAUTION

Even with alarms set, if “Not detected” has been set in the above alarm detection, no alarm is issued.

Be sure to set “Detected” in the above alarm detection if you use alarm features.