<8. Detailed Data Setting>

8-1

IM 11M12A01-04E

8. Detailed Data Setting

8.1 Setting Display Item

Display items are defined as items displayed on the basic panel display.
Parameter code

「A00」 or 「F08」 is used to set the display items as shown in Table 8.1. The oxygen

concentration is set at the factory before shipment. In addition, if the data initialization is performed,

the oxygen concentration will be set.

Table 8.1

Display Item

Values set with A00 or F08

Items displayed on the basic panel display

0

Indicates the oxygen concentration.

1 or 2

For humidity analyzers only. (if 1 or 2 is set for the oxygen analyzer,

"0.0" is only displayed on the basic panel display.)

3

Displays an item for the current output.

If the output damping has been set for the current output, values

involving the output damping are displayed.

CAUTION

If you set “3” in the parameter code

「A00」 or 「F08」, be sure to select “Oxygen Concentration” in the

following mA output setting (see Section 8.2, “Current Output Setting”).

8.2 Current Output Setting

This section describes setting of the analog output range. Table 8.2 shows parameter codes for the

set items.

Table 8.2

Current Output Parameter Codes

Set item

Parameter code

Set value

Analog output

C01

0 Oxygen concentration
1 4 mA (fixed *1)
2 4 mA (fixed *1)

Output mode

C03

0 Linear
1 Logarithm

Min. oxygen concentration

C11

Oxygen concentration at 4 mA

Max. oxygen concentration

C12

Oxygen concentration at 20 mA

Output damping constant

C30

0 to 255 seconds

*1:

For the oxygen analyzer, set 0 (zero) only for parameter code C01. When it is set, the current output is 4-mA fixed regardless of

the oxygen concentration.