<5. Wiring>

5-2

IM 11M12A01-04E

5.1.1 Terminals for the External Wiring

Remove the terminal cover on the opposite side of the display to gain access to the external wiring

terminals.

1 2 C

DO

DO

DI

DI

1

2

L N

FG

AO

G FG

F5101.ai

+

Figure 5.1

Terminals for External Wiring

5.1.2 Wiring

Make the following wiring for the equipment. It requires a maximum of four wiring connections as

shown below.

(1) Analog output signal

(2) Power and ground

(3) Contact output

(4) Contact input

1

DI-1

2

DI-2

3

DI-C

4

DO-1

5

DO-1

6

DO-2

7

DO-2

8

FG

9

AO

(+)

10

AO

(-)

11

L

12

N

13

G

14

FG

F28.EPS

Model ZR202G Integrated type Zirconia Oxygen Analyzer

Contact input 1

Contact output 2

Contact output 1

Analog output

4-20 mA DC

100 to 240 V AC,

50 or 60 Hz

Digital output

Contact input 2

The protective grounding for the analyzer shall be connected either the protective ground terminal

in the equipment or the ground terminal on the case.

Standard regarding grounding: Ground to earth, ground resistance: 100Ω or less.

Figure 5.2

Wiring Connection