Warning – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 127

Advertising
background image

<10. Other Functions>

10-18

IM 11M12A01-04E

(2) Next, adjust the flow rate to 600 ± 60 ml/min using the span gas valve “AIR” (the flow check

ball stops floating on the green line when the valve is slowly opened). To rotate the valve

shaft, loosen the lock nut and turn it using a flat-blade screwdriver. Turning the valve shaft

counterclockwise increases the flow rate.

(3) After adjusting the flow rate, tighten the valve lock nut.

(4) After the measured oxygen concentration is stabilized, touch the [ENT] key, then all the digits

flash. Touch the [ENT] key again to display “ZEro Y”. Disconnect the power cord to stop the

pump.

<Flow of zero gas>

Touch the [ENT] key to display a zero gas value set with the parameter code B01. Touch the [ENT]

key again to flash “OPEn” and the “measured oxygen concentration” alternately. To cause the zero

gas flow, follow these steps:

(1) Use the needle of the zero gas valve “CHECK GAS” to puncture a hole in the gas cylinder

installed as described in Section 10.5.2. Fully clockwise turn the valve regulator by hand.

(2) Next, adjust the flow rate to 600 ± 60 ml/min (the flow check ball stops floating on the green

line when the valve is slowly opened). Turn the regulator of the zero gas valve back slowly

counterclockwise. At that time, the flow rate also decreases as the inner pressure of the gas

cylinder decreases. Monitor the flow check and, when the ball’s position changes greatly,

readjust the valve.

(3) Touch the [ENT] key after the measured oxygen concentration becomes stable. Then all the

digits flash. Touch the [ENT] key again so that the “CALEnd” flashes.

NOTE

Be sure not to terminate the calibration in progress because of a shortage of gas in the cylinder. Each

gas cylinder is operable for nine minutes or more provided the gas is discharged at the specified rate.

Therefore, if your calibration time is estimated at four minutes, you can operate the zero calibration

twice.

(4) Stop the zero gas flow. Turn the zero gas valve regulator fully clockwise. If this valve regulator

is not properly adjusted, the needle valve will not close completely and a cylinder gas may leak.

When the output stabilization time elapses, the calibration is complete.

<Treatment after completion of calibration>

(1) Fully close the stop valve mounted on the calibration gas inlet of the detector.

(2) Remove the tube connecting the detector to the standard gas unit.

WARNING

Store the standard gas unit with the gas cylinder mounted where the ambient temperature does not

exceed 40°C. Otherwise, the gas cylinder may explode. Store the spare gas cylinders under the same

condition.