<3. Installation>

3-9

IM 11M12A01-04E

3.4 Installation of the Case Assembly (E7044KF) for

Calibration Gas Cylinder

The case assembly is used to store the G7001ZC zero gas cylinders.

3.4.1 Installation Location

The following should be taken into consideration:

(1) Easy access for cylinder replacement

(2) Easy access for checking

(3) Near to the detector and converter as well as the flow setting unit.

(4) The temperature of the case should not exceed 40°C due to rays of the sun or radiated heat.

(5) No vibration

3.4.2 Mounting

Mount case assembly on a pipe (nominal JIS 50 A) as follows:

(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting

the case assembly. (The sum of the case assembly and the calibration gas cylinder weighs

approximately 4.2 kg.)

(2) Mount the case assembly on the pipe by tightening the nuts with the U-bolt so that the metal

fitting is firmly attached to the pipe.

F3-16E.ai

A pipe to be mounted

(nominal JIS 50A : O.D. 60.5 mm)

Figure 3.16

Pipe Mounting