<3. Installation>

3-2

IM 11M12A01-04E

3.1.2 Probe Insertion Hole

CAUTION

• The outside dimension of detector may vary depending on its options. Use a pipe that is large

enough for the detector. Refer to Figure 3.1 for the dimensions.

If the detector is mounted horizontally, the calibration gas inlet and reference gas inlet should

face downwards.

• When using the detector with pressure compensation, ensure that the flange gasket does not

block the reference gas outlet on the detector flange. If the flange gasket blocks the outlet, the

detector cannot conduct pressure compensation. Where necessary, make a notch on the flange

gasket.

• The sensor (zirconia cell) at the probe tip may deteriorate due to thermal shock if water drops are

allowed to fall on it, as it is always at high temperature.

(1) Do not mount the probe with the tip higher than the probe base.

(2) If the probe length is 2.5 m or more, the detector should be mounted vertically (no more than a 5°

tilt).

(3) The detector probe should be mounted at right angles to the sample gas flow or the probe tip

should point downstream.

100mm

(vertical)

Bounds of the probe

insertion hole location

Flange matches

the detector size

(horizontal)

100mm

Four-hole flange

Eight-hole flange

F3101E.ai

*

1

Note

*1

Type

Outside diameter of detector

With dust filter

51 mm in diameter (Note)

With probe protector 60.5 mm in diameter (Note)

Standard

50.8 mm in diameter (Note)

(Note) When using the detector with pressure compensation, ensure that the flange gasket does not block the reference gas

outlet on the detector flange. If the flange gasket blocks the outlet, the detector cannot perform pressure compensation.

Where necessary, make a notch in the flange gasket.

Figure 3.1

Illustrates an example of the probe insertion hole

3.1.3 Installation of the Probe

CAUTION

• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop the detector,

as impact will damage it.

• A gasket should be used between the flanges to prevent gas leakage. The gasket material

should be heatproof and corrosion-proof, suited to the characteristics of the sample gas.