<8. Detailed Data Setting>

8-11

IM 11M12A01-04E

8.6 Input Contact Settings

The converter input contacts execute set functions by accepting a remote (contact) signal. Table 8.12

shows the functions executed by a remote contact signal.

Table 8.12

Input Contact Functions

Set item

Function

Calibration gas pressure

decrease

While a contact signal is on, neither semi-automatic nor automatic calibrations

can be made.

Measuring range change While contact input is on, the analog output range is switched to 0-25% O

2

.

Calibration start

If a contact signal is applied, semi-automatic calibration starts (only if the semi-

automatic or automatic mode has been setup). Contact signal must be applied

for at least one second. Even though a continuous contact signal is applied, a

second calibration cannot be made. If you want to make a second calibration,

turn the contact signal off and then back on.

Unburnt gas detection

If a contact signal is on, the heater power will be switched off. (An one-to 11-

second time interval single-output signal is available as a contact signal.) If this

operation starts, the sensor temperature decreases and an error occurs. To

restore it to normal, turn the power off and then back on, or reset the analyzer.

CAUTION

• Measurement range switching function by an external contact input is available for analog output

1 only and the range is fixed to 0-25%O

2

.

• To conduct a semi-automatic calibration, be sure to set the Calibration setup mode to “Semi-

automatic” or “Automatic”.

8.6.1 Setting Input Contact

To set the input contacts, follow the parameter codes given in Table 8.13.

Table 8.13

Parameter Codes for Input Contact Settings

Set item

Parameter code

Set value

Input contact 1 (function)

E01

0 Invalid
1 Calibration gas pressure decrease
2 Measuring range change
3 Calibration
4 Unburnt gas detection

Input contact 2 (function)

E02

0 Invalid
1 Calibration gas pressure decrease
2 Measuring range change
3 Calibration
4 Unburnt gas detection

Input contact 1 (action)

E03

0 Operated when closed
1 Operated when open

Input contact 2 (action)

E04

0 Operated when closed
1 Operated when open

8.6.2 Default Values

When the analyzer is delivered, or if data are initialized, the input contacts are all open.