<1. Overview>

1-2

IM 11M12A01-04E

CAUTION

• As this system uses ambient air for the reference gas, measuring accuracy will be affected by

the installation location.

• A stop valve should be connected to the calibration gas inlet of the equipment. The valve should

be fully closed unless calibration is in progress.

1.1.2 System 2

This system is for monitoring and controlling oxygen concentration in the combustion gases of a

large-size boiler or heating furnace. Instrument air (clean and dry air of oxygen concentration 21%)

is used as the reference gas and the span gas for calibration. Zero gas is fed from a cylinder during

calibration. The gas flow is controlled by the ZA8F flow setting unit (for manual valve operation).

~

ZR202G Integrated type Zirconia Oxygen Analyzer

F1.2E.ai

ZA8F flow setting unit

Reference gas

Calibration gas

Needle

valve

Flowmeter

Instrument air

Air Set

Calibration

gas pressure

regulator

Zero gas cylinder

Calibration gas unit case

Stop valve

or

Check valve

Span gas(Same as Zero gas calibration)

100 to 240 V AC

Contact input
Analog output, contact output

Digital output (HART)

Figure 1.2

Example of System 2