Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 23

Advertising
background image

<2. Specifications>

2-4

IM 11M12A01-04E

Functions

Display Function:

Displays values of the measured oxygen concentration, etc.

Alarm, Error Display:

Displays alarms such as “AL-06” or errors such as “Err-01” when any such

status occurs.

Calibration Functions: Auto-calibration; Requires the Automatic Calibration Unit. It calibrates

automatically at specified intervals.

Semi-auto Calibration; Requires the Automatic Calibration Unit. Input

calibration start signal by optical switch or contact, then it calibrates

automatically afterwards.

Manual Calibration; Calibration with opening/closing the valve of calibration

gas in operation interactively with the optical switch.

Maintenance Functions: Can operate updated data settings in daily operation and checking. Display

data settings, calibration data settings, test settings (current output loop

check, input/output contact check).

Setup Functions:

Initial settings suit for the plant conditions when installing the converter.

Current output data settings, alarm data settings, contact data settings, other

settings.

Display and setting content:

Display Related Items: Oxygen concentration (vol% O

2

), Output current value (mA), air ratio,

moisture quantity (in hot gases) (vol% H

2

O), Cell temperature (°C ),

thermocouple reference junction temperature (°C ), maximum/minimum/

average oxygen concentration (vol% O

2

), cell e.m.f. (mV), cell internal

resistance (Ω), cell condition (in four grades), heater on-time rate (%),

calibration record (ten times), time (year/month/day/hour/minute)

Calibration Setting Items: Span gas concentration (vol% O

2

), zero gas concentration (vol%O

2

),

calibration mode (auto, semi-auto, manual), calibration type and method

(zero-span calibration, zero calibration only, span calibration only),

stabilization time (min.sec), calibration time (min.sec), calibration period

(day/hour), starting time (year/month/day/hour/minute)

Output Related Items: Analog output/output mode selection, output conditions when warming-up/

maintenance/calibrating/abnormal, 4 mA/20 mA point oxygen concentration

(vol% O

2

), time constant, preset values when warming-up/maintenance/

calibrating/abnormal, output preset values on abnormal

Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol% O

2

),

Oxygen concentration low alarm/low-low alarm limit values (vol% O

2

),

Oxygen concentration alarm hysteresis (vol% O

2

), Oxygen concentration

alarm detection, alarm delay (seconds)

Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 and

2 (abnormal, high-high alarm, high alarm, low alarm, low-low alarm,

maintenance, calibrating, range switching, warming-up, calibration gas

pressure decrease, flameout gas detection (answer-back of contact input)