B1 0, Span y, Open/20.84 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 87: Zero y, Open/0.89, Calend

Advertising
background image

<8. Detailed Data Setting>

8-4

IM 11M12A01-04E

Switch operation

Display

>

ENT

b

1

0

>

ENT

CAL

>

ENT

CAL

>

ENT

SPAn

Y

>

ENT

21.00

%

>

ENT

OPEn/20.84

>

ENT

20.84

%

>

ENT

ZEro

Y

>

ENT

0.98

%

>

ENT

OPEn/0.89

>

ENT

0.89

%

>

ENT

CALEnd

>

ENT

b

10

>

ENT

Measured-value

display

Figure 8.1

Definition of During Calibration

In a semi-automatic calibration, “during calibration” is the time, starting when a calibration

instruction is executed with an infrared switch or a contact input, to make a series of calibrations,

until the preset output stabilization time elapses.

In an automatic calibration, “during calibration” is the time, starting when automatic calibration is

carried out at the calibration start time, until the preset output stabilization time elapses.

(4) “Error” appears when Error 1 to Error 4 are being issued

Output hold time

during calibration

Output hold time

during calibration