Installation of zr20h automatic calibration unit, Piping – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 54

Advertising
background image

<4. Piping>

4-5

IM 11M12A01-04E

• Installation of ZR20H Automatic Calibration Unit

Zero gas inlet
Rc1/4 or 1/4NPT(Female)

Reference gas inlet
Rc1/4 or 1/4NPT(Female)

Terminal box side

Span gas inlet
Rc1/4 or 1/4NPT(Female)

166.5

258

243

214

MAX

44

66.5

40

40

Vertical mounting on the ZR202G (-B)

F3.15E.ai

Horizontal mounting on the ZR202G (-A)

Unit: mm

Display side

Zero gas inlet
Rc1/4 or 1/4NPT(Female)

Span gas inlet
Rc1/4 or 1/4NPT(Female)

Reference gas inlet
Rc1/4 or 1/4NPT(Female)

166.5

60

45

160

66.5

40

40

180

MAX

44

Piping Diagram

SPAN IN

REF. IN

ZERO IN

F4.11E.ai

Calibration gas

Reference gas

ZR202G body

Span gas

solenoid valve

Span gas flowmeter

Reference gas flowmeter

To Air set

To Zero gas cylinder

Needle valve

Automatic Calibration Unit

Zero gas flowmeter

Zero gas solenoid valve

Check

valve