<11.Inspection and Maintenance>

11-4

IM 11M12A01-04E

F11.3E.ai

Dust filter

(optional)

Bolts (four)

Washers

(four)

Sensor (cell)

U-shaped pipe

Metal O-ring

Contact

Probe

1/8 turn – tighten bolts 1/8 turn

(approximately 45°) each

U-shaped pipe

support

Screw

Filter

Figure 11.3

Exploded View of Sensor Assembly

CAUTION

Optional Inconel bolts have a high coefficient of expansion. If excess torque is applied while the bolts

are being tightened, abnormal strain or bolt breakage may result. So, tighten the bolts following the

instructions given above.

11.1.3 Replacement of the Heater Assembly

This section describes the replacement procedure for the heater assembly.
The sensor or ceramic heater-furnace core internal structure is subject to fracturing, so do NOT

subject it to strong vibrations or shock. Additionally, the heater assembly reaches high temperatures

and is subjected to high voltages.
So, maintenance services should be performed after the power is off and the heater assembly

temperature has returned to normal room temperature.
For details, refer to IM11M12A01-21E “ Heater Assembly “.

NOTE

If the heater assembly can not be removed because a screw has fused to its thread, one of our

service representatives can fix it.