1 contact output check, Contact output check -12, G0 1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 77: G1 1, Caution

Advertising
background image

<7. Startup>

7-12

IM 11M12A01-04E

7.9.1 Contact Output Check

Follow Table 7.8 to check the contact output. The table uses an example with contact output 1.

Table 7.8

Checking Contact Output

Switch operation

Display

Description

>

ENT

A

01

Display after the password has been entered.

>

ENT

G

01

Touch the [∧] key to switch to Group G.

>

ENT

G

0

1

Touch the [>] key to move the position of the digit that is flashing to the right one
digit.

>

ENT

G

1

1

Touch the [∧] key to enter 1.

>

ENT

0

Touch the [ENT] key to have 0 (zero) flash. The contact is then open.

>

ENT

1

Touch the [∧] key to set 1 (one).

>

ENT

1

Touch the [ENT] key. The flashing continues.

>

ENT

1

Touch the [ENT] key again to stop the flashing, and the contact will be closed.

>

ENT

G

1

1

Touch the [ENT] key once again to switch to the parameter code selection display.
The contact then returns to the original state.

>

ENT

Basic

panel

display

Touch the [>] key together with the [ENT] key to return to the basic panel display.
(This is not required if you proceed to make another setting.)
(The displayed numeric characters indicate the measurement gas concentration.)

The symbol [

] indicates that the key is being touched. Light characters indicates that the digits are flashing.

CAUTION

• If you conduct an open-close check for the contact output 2, Error 1 (cell voltage failure) or Error

2 (heater temperature abnormal) will occur. This is because the built-in heater power of the

detector, which is connected to contact output 2, is turned off during the above check. So, if the

above error occurs, reset the equipment or turn the power off and then back on to restart (refer to

Section 10.4, “Reset,” later in this manual).