<4. Display and Operations while Connected to GC8000>

4-57

IM 11B06C01-01E

l

Displaying retention time (history) results from the past

1) Select Calibration Coefficient History Window in the Window menu, or the Calibration

Coefficient History Window command button on the toolbar to display the Calibration

Coefficient History window.

2) Select SYS Number Select in the Display Item menu, or the system number or GCM of the

retention time result you want to display from the system number select combo box on the

toolbar.

3) Select Open History in the File menu, or the Open History command button on the toolbar to

open the history setting dialog box.

4) From the radio button selecting Display from Latest or Display from History, select Display

from History, then apply it by pressing the OK button.

5) From the Display from History combo box, select the desired year, month and day, and

apply it using the OK button.

Figure 4.71

Year and month setting dialog box

6) The calibration coefficient results are displayed for up to 5,000 past analyses tracing back

from the specified file. The analysis results are then not automatically updated.

5th Edition : Nov. 16, 2011-00