1 toolbar, 2 gcm status display area, Toolbar -4 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 51: Gcm status display area -4

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-4

IM 11B06C01-01E

4.2.1 Toolbar

The toolbar consists of the toolbar buttons and the text window to indicate the ASET status.

Window Display button

Analyzer Operation window

Analysis Results windows

Chromatogram window

Alarm windows

Analyzer Status window

Change GCM

No. button

Change User

Level button

GC8000 Alarm

GCM No.

User Level

Change All VGC

Operation Mode button

Help button

Figure 4.3

Detailed explanation of toolbar

Table List of each element functions

Button/Display

Function and Display Information

Change User Level button Displays the Change User Level dialog box.

Use this button to change the user level (A, B, C, C+, or D).

Window Display button

Displays the relevant window. If a window cannot be displayed, its button is grayed

out.

Change GCM No. button

Displays the Change GCM Number dialog box. Use this button to change the

active GCM.

Change All GCM

Operation Mode button

Displays the Change All GCM Operation Mode dialog box.

This button changes the operation mode of all GCMs of the analyzer

simultaneously.

Help button

Displays a Help window.

GC8000 Alarm

Displays the alarm status of the entire GC8000. The most serious alarm status

level of each GCM is indicated.

l

Green: No alarm

l

Yellow: Level-2 alarm has occurred.

l

Red: Level-1 alarm has occurred.

The alarms have the following severity levels. Level 1 is the highest level of

severity. Level 1 → Level 2 → (Level 3) → No alarm

GCM#

Indicates the active GCM number.

This is empty (deselected) when this window is opened.

User Level

Displays the currently set user level.

4.2.2 GCM Status Display Area

The GCM Status Display area shows each GCM status of the GC8000. The number of display

areas depends on the GCM count (1 to 6) held by the GC8000.

GCM Alarm Status

GCM No.

Operation

Mode

Measurement Status

System No.

Elapsed time/System

Analysis cycle

Elapsed time

bar graph

Stream No.

Measurement status

Method No.

Changed Operation Mode Number

Display Analysis

Result button

Figure 4.4

Detailed description of GCM Status Display area