15 pressure display, Pressure display -18, Caution – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 65

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-18

IM 11B06C01-01E

4.3.15

Pressure Display

Display the pressure of each EPC (Electric Pressure Controller) in the following procedure. You

display the pressure of Oven 1 in this example.
From the Manual/Range menu, select “Oven1” and “Press1” commands. The Pressure Display

dialog box for Oven 1 appears.

Figure 4.22

Pressure Display dialog box

The Pressure Display dialog box shows the SV value, PV value, On/Off status and pressure

status of Carriers n-1 and n-2 and Utilities n-1 to n-4.
The display is refreshed in every second. The pressure status is Under preparation, Constant

pressure, or Pressure increasing.
You can also call the Pressure Display dialog box by clicking the pressure display of the oven.

You cannot start an operation from this dialog box. Also, carriers may not be used except for

Carrier n-1. In such case, the corresponding items are left blank.

CAUTION

This pressure display is valid only when the EPC (Electric Pressure Controller) is used.