Aset for arcnet, 1 system configuration, Aset for arcnet -1 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 177: System configuration -1

Advertising
background image

<8. ASET for ARCNET>

8-1

IM 11B06C01-01E

8.

ASET for ARCNET

The New ARCNET-based bus system enables connection between an ARCNET system

and an Ethernet system. This realizes addition of Ethernet analyzers to existing ARCNET

systems.

8.1

System Configuration

ASET-G software which is ARCNET supported version must be installed to corresponde with the

ARCNET system.
A typical network configuration of analyzer bus is shown in Figure 8.1.
The network consists of analyzers (Gas chromatograph etc.) and "Analyzer bus systems for

ARCNET". This "Analyzer bus systems for ARCNET" consist of “Gateway unit”, ”analyzer server”

and “ARCNET / Ethernet Converter”.
Each software (PCAS, ASET) for “Analyzer server” and “Analyzer Server Engineering Terminal”

should be used “/ARC: ARCNET supported version”.
The communication to the analyzer connected to Ethernet is also enabled.

• "Analyzer bus systems for ARCNET" should be used so that DCS communicates with

analyzer connected to ARCNET.

• Redundancy for ARCNET is enabled. Redundancy for Ethernet can be done by using

double "Analyzer bus systems for ARCNET".

• One "Analyzer bus systems for ARCNET" can communicate with up to 30 sets of analyzers

connected to ARCNET. In case of communicating with more than 31 sets, plural number of

"Analyzer bus systems for ARCNET" should be used.

F8-0.ai

DCS

DI/O

RS-232C

LAN (Ethernet)

Analyzer bus system

for ARCNET

ARCNET

PC

[PCAS/ARC]

PC

[ASET/ARC]

STARDOM FCJ

[ASGW]

ASIU

GCIU

GC1000 MarkII

GC1000

GC1000

[ ]:Software

ARCNET/Ethernet

Converter

GC1000 MarkII

GCIU

Analyzer

pH analyzer,

Oxygen analyzer, etc.

Network monitoring,

Automatic data storage

Gateway

Operation and display of

Gas Chromatograph

Figure 8.1

Typical System configuration

5th Edition : Nov. 16, 2011-00