3.7

Uploading and Downloading Analyzer

Setting Parameters

When incorporating the settings specified in the Set Parameter List window of the LCD panel

into another analyzer, or when creating a backup for re-setting, you upload and download the

parameters.
Before transferring the parameters, you need to specify the ID number of the analyzer to operate.

SEE ALSO

For specifying the analyzer ID number, see section 3.6, “Selecting Analyzer.”

CAUTION

Only level-C users can upload and/or download setting parameters. Also, the operation status

must be changed to “Stop.”

n

Uploading

To upload the set parameters from the analyzer and save them to the Engineering Terminal disk,

proceed as follows:

l

Operation procedure

(1) Click on the Parameter Transfer command in the Option menu.

The submenu for transferring parameters appears.

F0321.ai

(2) Click on the Upload to PC command.

The Save As dialog box for specifying a setting parameter file appears.

F0322.ai

Enter a file name.

5th Edition : Nov. 16, 2011-00