18 stream valve on/off setting, Stream valve on/off setting -21 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 68

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-21

IM 11B06C01-01E

Table Display of Range Change dialog box for each process setting (details)

Process setting

(Details)

Result unit

Measuring

range

Reference peak

No. 1

Reference peak

No. 2

Remarks

External third

Displayed

Displayed

Displayed

Negative (-) sign

is shown if the

reference stream

and peak are “0,

0”.

Negative (-) sign is

shown.

For Calibration

Stream, the

reference stream

and peak are “0,

0”.

External linear

Displayed

Displayed

Displayed

Displayed

For Calibration

Stream, the

reference stream

and peak (for both

1 and 2) are “0, 0”.

Indirect method

Deviation

calculation

Linear 1 to 5

Displayed

Displayed

Negative (-) sign is

shown.

Negative (-) sign is

shown.

AI1 to 16

Negative (-)

sign is shown.

Displayed

0% value

100% value

Others

(such as ratio)

Negative (-)

sign is shown.

Displayed

Negative (-) sign is

shown.

Negative (-) sign is

shown.

4.3.18

Stream Valve ON/OFF Setting

To set a turn-On/Off time of stream valves, select the Stream Valve command from the Setup

menu. The Stream Valve ON/OFF Setting dialog box appears.

Figure 4.25

Stream Valve ON/OFF Setting dialog box

If no stream valve is used for the active GCM, the Stream Valve command is grayed out on the

Setup menu and you cannot select it.
In this dialog box, you can change the ON and OFF time for each method number of stream

valve of the active GCM. First, select a method number. A pair of ON and OFF time of the stream

valve of this method is displayed. Change the ON and OFF times and click the Change button.

A confirmation message asking “Are you sure you want to send the data?” is displayed in the

message box. When you respond with “Yes”, the settings are changed.
Each item can be selected in the following range.
Method number: 1 to 6
Valve ON time: “- Preparation time to System analysis cycle - 2 sec”
Valve OFF time: “ON time to System analysis cycle - 2 sec”
However, if it is set to “********”, both ON and OFF times are made invalid.

5th Edition : Nov. 16, 2011-00