<4. Display and Operations while Connected to GC8000>

4-38

IM 11B06C01-01E

l

Chromatogram re-integration area

Re-integration button

Individual peak data area

Analysis result list

Chromatogram information

Figure 4.44

Chromatogram re-integration area (external third)

1) Chromatogram information

Displays the analysis start time, stream number, and detector number of chromatogram to

be re-integrated.

The location of the parameter file can also be specified. If it is acquired from GC8000 while it

is currently communicating, “Online” is displayed.

2) Analysis result list

Displays the list of analysis results corresponding to chromatogram to be re-integrated.

Displayed in the list are relative peak number, peak name (unit), concentration, and

retention time.

By clicking a peak in the list, you can select the peak to be re-integrated. The peak

information currently selected is highlighted in cyan.

3) Individual peak data area

Displays the current analysis results of the selected peak. Information required for re-

integration can be edited in this area.

Items to be displayed vary depending on the combination of the processing setting and the

concentration calculation setting.

4) Re-integration buttons

Re-integration button: Use this button to perform re-integration.

Update button: Use this button to apply the re-integration results. (The results are not saved

to a file)

Peak processing/transfer button: Use this button to transfer gate time and other data edited

in re-integration to the initial GC8000 database.

Result transfer button: Use this button to save the re-integration results to the chromatogram

file or analysis result file, and transfer them to PCAS or GC8000. However, the results

cannot be transferred unless you click the update button previously.

Cancel button: Use this button to discard the re-integration results and restore the status

before re-integration. If the update button was clicked previously, however, the status before

re-integration cannot be restored .