<4. Display and Operations while Connected to GC8000>

4-40

IM 11B06C01-01E

3) Specify the stream number, analysis start time, and the number of re-integrating

processes. Subsequently, clicking the “Execute” button performs re-integration of multiple

chromatograms. At this time, re-integration is performed for the analysis results of the

analysis count tracked back from the analysis start time.

4) When re-integration has been completed, the execution result (success or fail) of the

individual chromatogram re-integration is displayed in the dialog box. Failed re-integration

results are not updated.

5) To exit, click the “Close” button. Then, the Re-integrate Multiple Chromatogram dialog box

closes.

Multiple chromatogram re-integration is determined to fail in the following cases:

• The number of peak assignments is different from that in the re-integration reference

chromatogram.

• Peak detection failed.
• The concentration value after re-integration falls out of the range.

CAUTION

“Distillation = Yes” is set in the analyzer currently connected or loaded parameter file, re-

integration cannot be performed.

4.4.11

Pen Setting

Select the Set Pen command in the Graph menu to display the Set Pen dialog box.

Figure 4.45

Set Pen dialog box (Example in the right shows a combo box is selected)

From this dialog box, different types of chromatograms to be displayed with pen 1 to 8 can be

selected.
From a combo box for each pen color, select a chromatogram you want to display. Depending on

the type of selected chromatogram, a different dialog box may be invoked.
The following shows the graph names that can be selected from the combo box.