<5. Display and Operations while Connected to GC1000 Mark II>

5-15

IM 11B06C01-01E

l

Changing the measurement status

To change the measurement status, click the button of the measurement status you wish

to change to. The operable button(s) differs depending on current conditions. The operable

button(s) is displayed in black; any status(es) not acceptable is grayed out.
There is also a “Cancel Setting” button

. This is a designation-dedicated button that cancels

the queued measurement status change.

l

Calibration (validation) methods

There are three types of calibration (validation) methods as follows:

Calibration (Validation) Method

Description

Automatic

Automatically conducts calibration (validation) according to the start time

and the analyzer’s pre-set time period.

Semi-automatic

Click the Calibration (Validation) button, specify the execution number,

and touch the Run button to perform calibration (validation).

Manual

After clicking the Calibration (Validation) button, make preparations, then
using the Start Calibration (Validation) button

and the End Calibration

(Validation) button

conduct calibration (validation) manually.

Selection of a calibration (validation) method is made as follows:
(1) Click the Calibration (Validation) Method button

.

The Calibration (Validation) Method dialog box appears.

F5--2-14.ai

(2) Check the radio button of the calibration (validation) method you wish to specify, then click

the Execute button. A change confirmation dialog box appears.

F5-2-15.ai

(3) Click the Yes button.
(4) A message appears informing you that data transmission is complete.

F5-2-16.ai

(5) Click the OK button. This completes the selection of the calibration (validation) method.

5th Edition : Nov. 16, 2011-00