<4. Display and Operations while Connected to GC8000>

4-13

IM 11B06C01-01E

4.3.4

Changing Stream Sequence No.

To change a stream sequence number, select the Stream Sequence command from the

Measurement Status menu. When the Stream Sequence dialog box appears, select a sequence

number and click the Execute button. The measurement status is changed to the stream

sequence of the specified number. (The sequence number can be 1 to 8.)

TIP

The stream sequence number will not be changed if any stream has not been allocated to the specified stream

sequence number.

Figure 4.15

Stream Sequence dialog box

4.3.5

Changing Stream Specification

To change a stream specification, select the Stream Specification command from the

Measurement Status menu. The Stream Specification dialog box appears.

Only stream numbers

belonging to GCM

can be selected.

Figure 4.16

Stream Specification dialog box (The box when opened is shown at left, and the box for

stream number selection is shown at right.)

Select a stream number and the iteration count, and click the Execute button. The measurement

status is changed to the stream specification and the stream is measured for the specified

number of times. Then, the operation is stopped automatically.
If you set the iteration count to zero (0), the specified stream is measured continuously.

5th Edition : Nov. 16, 2011-00