6 latest analysis results window, Latest analysis results window -9 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 56

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-9

IM 11B06C01-01E

2) When you respond to the confirmation message with OK, the operation mode is changed

for each GCM.

If the operation mode change request is rejected due to the operation conditions, a level-3 alarm

is generated and its status information is given.
When you issue a Change All GCM Operation Mode request, multiple GCMs may not accept the

specified operation mode. In such case, multiple sets of level-3 alarm are generated.

4.2.6 Latest Analysis Results Window

The Latest Analysis Results window lists the latest concentration of selected GCMs.

Figure 4.11

Latest Analysis Results window

The stream number and peak name are shown at left, and the concentration is shown at right.

The concentration is shown as “???????” if it exceeds the upper or lower limit.
Up to 999 sets of peak concentration can be displayed. You can display a concentration outside

of the window page by scrolling it.
The display of this window is refreshed when the latest analysis result of each GCM is obtained.

5th Edition : Nov. 16, 2011-00